SPECIFICATIONS | ||
AC HI-POT SPECIFICATIONS | ||
Voltage Range | 0.100 ~ 5.000kV | |
Voltage Resolution | 5V | |
Voltage Regulation | 1% + 5 V | |
Voltage Accuracy | ±1% of reading ±10 V (above 500V) | |
Max. Current | 40mA, (above 500V) | |
Current Limit | 0.10 ~ 40mA, 0.02mA (resolution) | |
Current Accuracy | ±1% of reading ±80μA | |
DC HI-POT SPECIFICATIONS | ||
Voltage Range | 0.100 ~ 6.000kV | |
Voltage Resolution | 5V | |
Voltage Regulation | 1% + 5 V | |
Voltage Accuracy | ±1% of reading ±10 V (above 500V) | |
Max. Current | 10mA, (above 500V) | |
Current Limit | 0.10~10mA, 0.01mA (resolution) | |
Current Accuracy | ±1% of reading ±50μA | |
INSULATION RESISTANCE SPECIFICATIONS | ||
DC Voltage | 50V/100V/500V/1000V | |
Resistance Range | 1 ~ 9900M(500V,1000V); 1 ~ 2000M(50V,100V) | |
Resistance Accuracy | 50V/100V | |
1~50MΩ: ±5% of reading ±1Count | ||
51~200MΩ: ±10% of reading ±1Count | ||
201M~1990MΩ: ±20% of reading ±1Count | ||
500V/1000V | ||
1~500MΩ: ±5% of reading ±1Count | ||
501~2000MΩ: ±10% of reading ±1Count | ||
2001M~9900MΩ: ±20% of reading ±1Count | ||
GROUND BOND SPECIFICATIONS | ||
Voltage Range | Max. 8V (DC) | |
Current Range | 3 ~ 42A | |
Current Resolution | 0.02A | |
Current Accuracy | 1% of reading 0.2A(3A~8A) | |
Ground Bond | 1% of reading 50mA(9A~42A) | |
Resistance Test | 0 ~ 620mΩ | |
Resistance Resolution | 0 ~ 600mΩ(100μΩ) | |
Resistance Accuracy | ±1% of reading 3 mΩ | |
Test Method | Four terminals test mode | |
ARC DETECT | ||
Detect Current | 40 level (1 ~ 40mA) | |
INTERFACE | ||
Remote I/O | Standard | |
RS-232C | Standard | |
Scan | Standard | |
Real Panel Output | Standard | |
GPIB | Option | |
Scanner Box | Option (HSB-001-1 : 8ch HV+8ch GB ; HSB-001-2 : 16ch HV) | |
POWER SOURCE | AC 100V, 120V, 220V, 230V ± 10%, 50/60Hz | |
ACCESSORIES | ||
Instruction manual x 1 | ||
Power cord x 1 | ||
Test leads GHT-105A x 1, GTL-115 x 1 | ||
DIMENSIONS & WEIGHT | 330(W) x 149(H) x 446(D) mm | |
Approx.15kg |