2 Channels, 4 Traces: DC ~ 30MHz Bandwidth
1mV/div ~ 10V/div
CRT Readout
ALT MAG Function (x 5, x 10, x20)
Vertical Mode Trigering
TV Synchronization
CH1 Signal Output, Z-axis Input
Buzzer Alarm & LED Indicators
Compact size (275Wx130Hx370Dmm) and Front Panel Layout Groups for easy to use | |
|
CRT
|
Type |
6-inch rectangular type with internal graticule; 0%, 10%, 90%, 100% markers. 8x10 div(1div=1cm) |
Accelerating Potential |
Approx. 2kV |
Trace Rotation |
Provided |
Z-axis Input |
Sensitivity : at least 5V ; Polarity : positive going input decrease intensity Usable frequency range : DC to 2MHz ; Max. input voltage : 30V (DC + AC peak) at 1kHz or less Input Impedance : Approx.47kΩ |
VERTICAL SYSTEM
|
Sensitivity and Accuracy |
1mV ~ 2mV/div ±5%, 5mV ~ 20V/div ±3%, 14 calibrated steps in 1-2-5 sequence |
Vernier Vertical Sensitivity |
Continuously variable to 1/2.5 or less of panel indicate value |
Bandwidth (-3dB) and Rise Time |
|
Bandwidth (-3dB) |
Rise Time |
5mV ~ 20V/div
|
DC ~ 30MHz
|
Approx. 11.7nS |
1mV ~ 2mV/div |
DC ~ 7MHz |
Approx. 50nS | |
Maximum Input Voltage |
400V (DC + AC peak) at 1kHz or less |
Input Coupling |
AC, DC, GND |
Input Impedance |
1MΩ ±2%//approx. 25pF |
Vertical Modes |
CH1, CH2, DUAL (CHOP, ALT), ADD, CH2 INV |
Chop Frequency |
Approx. 250kHz |
Dynamic Range |
8 div at 20MHz, 6div at 30MHz |
HORIZONTAL SYSTEM
|
Sweep Time |
0.2μs/div ~ 0.5s/div, 20 steps selectable in 1-2-5 sequence, continuous variable control between steps at least 1 : 2.5 |
Accuracy |
±3%, ±5% at x5 / x10MAG, ±8% at x 20MAG |
Sweep Magnification |
x5, x10, x20 MAG |
Maximum Sweep Time (at MAG) |
50ns/div (10ns/div~40ns/div are uncalibrated) |
ALT-MAG Function |
Available |
TRIGGER
|
Trigger Mode |
AUTO, NORM, TV |
Trigger Source |
VERT-MODE, CH1, CH2, LINE, EXT |
Trigger Coupling |
AC, HFR, LFR, TV-V(-), TV-H(-) |
Trigger Slope |
“+”or“ - ”polarity |
Trigger Sensitivity |
|
20Hz~2MHz |
2MHz~20MHz |
20MHz~30MHz |
CH1, CH2 |
0.5div |
1.5div |
2.0div |
Vert-Mode |
2.0div |
3.0div |
3.5div |
EXT |
200mV |
800mV |
1V |
TV sync pulse more than 1 div (CH1, CH2, VERT-MODE) or 200mV (EXT) |
External Trigger Input |
Input impedance: Approx.1MΩ//25pF(AC coupling);Max. input voltage: 400V(DC+AC peak)at 1kHz |
Hold-off Time |
Variable |
X-Y OPERATION
|
Input |
X-axis : CH1 ; Y-axis : CH2 |
Sensitivity |
1mV/div ~ 20V/div |
Bandwidth |
X-axis : DC ~ 500kHz (-3dB) |
Phase Difference |
3°or less from DC to 50kHz |
OUTPUT SIGNAL
|
CH1 Signal Output |
Voltage : approx. 20mV/div (with 50Ω terminated) Bandwidth : 50Hz to at least 5MHz |
Calibrator Output |
Voltage : 0.5V±3% Frequency : approx. 1kHz, square wave |
CRT READOUT
|
Panel Setting Display
|
CH1/CH2 sensitivity, sweep time, trigger condition |
POWER SOURCE |
|
|
AC 100V/120V/230V±10%, 50/60Hz |
ACCESSORIES
|
|
Instruction manual x 1 Power cord x 1 GLF-190C Probes (10:1/1:1) x 2 |
DIMENSIONS & WEIGHT |
|
|
275(W) x 130(H) x 370(D) mm Approx. 7.2kg | |
Inquiry |
Model |
Name |
Descripation |
|
GTC-002 |
Instrument Cart |
30kg Loading, 120V Input, 330(W) x 430(D) mm For GDS,GOS,GRS Series |
|
GTC-001 |
Instrument Cart |
30kg Loading, 120V Input , 450(W) x 430(D)mm For GDS,GOS,GRS Series |
|
GLF-190C |
Probe |
10:1 DC~60MHz, 1:1 DC~6MHz For: GOS-652G/653G, GOS-6000 Series, GRS-6000A Series | | | |